PCI-DB.com
  1. Driver
  2. Firmware
  3. Thermo Fisher 48i-HL Analyzer Firmware 02.02.03.304

Thermo Fisher 48i-HL Analyzer Firmware 02.02.03.304 Download

Posted at March 28, 2024 by PCI-DB Team

Install Driver Automatically
Device NameThermo Fisher 48i-HL Analyzer Firmware 02.02.03.304
CategoryFirmware
ManufacturerThermo Fisher
File Size15.4 MB
Supported OSWindows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Thermo Fisher 48i-HL Analyzer Firmware 02.02.03.304 Description

Version 02.02.03 changes relative to version 02.02.00:

1. Fix display of drift compensation for 410.
2. Added concentration calibration data to output list for MODBUS usage.
3. Fix logged data averaging where the sum was divided by 'n + 1' instead of 'n' seconds, yielding a possible 1.6% error in logged concentration data for 1-minute logging or 0.03% error in hourly concentration data.
4. Fix handling of IP addresses with 3 digits in each of the four fields in INSTRUMENT CONTROLS> COMMUNICATIONS> TCP/IP SETTINGS> IP ADDRESS screen.
5. Fix "set format", "set lrec format", "set srec format", and "set erec format" CLINK commands to save the new setting if the source of the command is serial.
6. Fixed "set program bootloader procboard" CLINK command so that it no longer causes iSeries+ units to hang.
7. The "host name" CLINK command now responds with a <space> if no host name is found. If "set host name" data does not pass validation, it now responds with "data not valid" instead of "bad command".

UPGRADE FIRMWARE

Below is a procedure for loading the firmware into FLASH memory. The firmware update file transfer process should take about 30 minutes at 57,600 baud. It is assumed that iPort is already talking to the instrument and the instrument window is currently open.

NOTE:

DO NOT TURN OFF THE INSTRUMENT AT ANY TIME DURING THIS UPDATE If the instrument is turned off while burning the new image to the FLASH, it may require replacement of the CPU board, motherboard, I/O expansion board, and/or measurement interface board. To reduce this risk, make sure the instrument is running on clean and stable power before performing this update.

1. Close all instrument windows.
2. From the iPort menu, select Instrument > Update Firmware. Select TCP/IP or Serial, depending on the connection.
3. In the Update Instrument Firmware Program dialog box, enter the instrument ID (if using serial port) or the TCP/IP address (if using TCP/IP).
4. In the Open File dialog box, select the firmware update file, then click the Open button.
5. File transfer progress can be monitored by looking at the transferred blocks in the lower left corner of the iPort window as well as on the instrument's display.
6. Once the file transfer is complete, the instrument will automatically reboot. There may be some error messages regarding configuration and calibration files that are displayed, this is normal after a firmware update. At this time, the bootloader and application code in each of the low-level processors will be updated to the latest version.
7. To verify all updates were successful, go to the ALARMS menu and make sure the board status alarms at the bottom of the menu all show “OK”. If any board status alarms show “FAIL”, try rebooting the instrument and checking the ALARMS menu again. If they still show “FAIL”, contact technical service.

About Device Firmware:

Applying a newer firmware version than the one already installed on your unit can bring various enhancements, include workarounds for diverse problems encountered by the device, and improve or add newly developed features.

On the other hand, downgrading the firmware version can recover the device’s functionality in the event of a software update either turning up faulty or causing the unit’s overall performance to drop. However, bear in mind that applying an older firmware isn’t recommended and might not always be possible.

When it comes to the update steps, due to the large variety of devices and numerous ways for installing a new firmware, it is recommended that you carefully read and understand the upgrade steps, and consider applying a different firmware only when you have familiarized yourself with the update method.

As the installation process is most of the times quite risky, this step should be taken by users that have the ability to successfully complete the update; regular users may initiate it at their own risk. Moreover, it’s best that this task be performed in a steady power environment such as the one ensured by a UPS unit.

Therefore, if you consider applying this release, hit the download button, take into account all the aforementioned aspects, and update the device’s firmware. Also make sure you constantly check with our website to ensure that you don’t miss a single new release. 

  It is highly recommended to always use the most recent driver version available.

Try to set a system restore point before installing a device driver. This will help if you installed an incorrect or mismatched driver. Problems can arise when your hardware device is too old or not supported any longer.

Related Thermo Fisher Drivers

Thermo Fisher 15i Analyzer Firmware 02.02.03.304

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 55i Analyzer Firmware 02.02.02.190

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 48i-HLJ Analyzer Firmware 02.02.03.304

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 81i Analyzer Firmware 02.02.09.174

Device Type: Firmware File Size: 15.3 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 48i-TLE Analyzer Firmware 02.02.03.304

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 60i Analyzer Firmware 02.02.08.348

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 46i Analyzer Firmware 02.02.03.304

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 48i-HL Analyzer Firmware 02.02.03.304

Device Type: Firmware File Size: 15.4 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Thermo Fisher 146i Analyzer Firmware 02.02.02.309

Device Type: Firmware File Size: 15.5 MB Windows XP ,Windows Vista ,Windows XP 64 bit ,Windows Vista 64 bit ,Windows 7 ,Windows 7 64 bit ,Windows 8 ,Windows 8 64 bit ,Windows 8.1 ,Windows 8.1

Install Driver

Find Missing Drivers

© 2024 PCI-DB.com - PCI Database Replacement. All rights reserved.